Ultra-low iodine concentrations iso-attenuating with diagnostic 0.5 M gadolinium in endovascular procedures to minimize the risk of contrast nephropathy: A phantom study
- 1. Department of Radiology, Oncology and Radiation Sciences, Section of Radiology, Uppsala University Hospital, SE-751 85 Uppsala (Sweden)
- 2. Department of Medical Physics, Uppsala University Hospital, SE-751 85 Uppsala (Sweden)
- 3. Institution of Clinical Sciences, Lund University, SE-221 00 Lund (Sweden)
Description
Purpose: To establish the concentrations of iodine contrast media (I-CM) iso-attenuating with 0.5 M gadolinium contrast media (Gd-CM), regarded diagnostic in catheter angiography and vascular interventions in azotemic patients, at various X-ray tube potentials with correlation to skin radiation dose. Materials and method: 20-mL syringes filled with 30, 40, 50, 70 and 90 mg I/mL, 0.5 M Gd-CM and air were placed in a water-equivalent phantom and exposed at about 50, 60, 70, 80 and 90 kV in an X-ray angiographic system. Relative contrast between the contrast materials and the background phantom material was measured on a PACS workstation. Radiation entrance dose, measured with a dose meter and estimated from the dose-area-product (DAP), was adjusted for radiation backscatter to simulate absorbed skin dose. Result: The iodine concentrations 30, 40, 50, 70 and 90 mg/mL resulted in the same relative contrast as 0.5 M gadolinium at 53, 57, 62, 71 and 85 kVp, respectively. Air had lower relative contrast than all iodine concentrations at all kVp-settings except for 30 mg I/mL above 84 kVp. The measured skin radiation dose was less than 1 mGy per exposure at all kVp-settings, and around 25–30% lower than the dose estimations derived from the angiographic system's in-built DAP meter. Conclusion: Low-kilovoltage X-ray technique and ultra-low concentrations of I-CM iso-attenuating with 0.5 M Gd-CM may be utilizable in peripheral arteriography and endovascular interventions, to minimize the total CM-dose to avoid CIN in azotemic patients
Availability note (English)
Available from http://dx.doi.org/10.1016/j.ejrad.2015.03.014Additional details
Identifiers
- DOI
- 10.1016/j.ejrad.2015.03.014;
- PII
- S0720-048X(15)00136-9;
Publishing Information
- Journal Title
- European Journal of Radiology
- Journal Volume
- 84
- Journal Issue
- 6
- Journal Page Range
- p. 1068-1074
- ISSN
- 0720-048X
- CODEN
- EJRADR
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 47049153
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ABUNDANCE; BIOMEDICAL RADIOGRAPHY; BLOOD VESSELS; CONCENTRATION RATIO; CONTRAST MEDIA; FAILURES; GADOLINIUM; HAZARDS; IODINE; KIDNEYS; PATIENTS; PHANTOMS; RADIATION DOSES; SKIN; X-RAY TUBES
- Descriptors DEC
- BODY; CARDIOVASCULAR SYSTEM; DIAGNOSTIC TECHNIQUES; DIMENSIONLESS NUMBERS; DOSES; ELECTRON TUBES; ELEMENTS; EQUIPMENT; HALOGENS; MEDICINE; METALS; MOCKUP; NONMETALS; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; RARE EARTHS; STRUCTURAL MODELS; X-RAY EQUIPMENT