Published August 2006
| Version v1
Journal article
Heterodyne interferometer with unequal path lengths
Creators
- 1. Applied Physics, California Institute of Technology, Pasadena, California 91125 (United States)
Description
Laser interferometry is an extensively used diagnostic for plasma experiments. Existing plasma interferometers are designed on the presumption that the scene and reference beam path lengths have to be equal, a requirement that is costly in both the number of optical components and the alignment complexity. It is shown here that having equal path lengths is not necessary, instead, what is required is that the path length difference be an even multiple of the laser cavity length. This assertion has been verified in a heterodyne laser interferometer that measures typical line-average densities of ∼1021/m2 with an error of ∼1019/m2
Additional details
Identifiers
- DOI
- 10.1063/1.2336769;
- arXiv
- arXiv:physics/0608082v1;
Publishing Information
- Journal Title
- Review of Scientific Instruments
- Journal Volume
- 77
- Journal Issue
- 8
- Journal Page Range
- p. 083503-083503.6
- ISSN
- 0034-6748
- CODEN
- RSINAK
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 38026496
- Subject category
- S70: PLASMA PHYSICS AND FUSION TECHNOLOGY; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ALIGNMENT; ERRORS; INTERFEROMETERS; INTERFEROMETRY; LASER CAVITIES; LASERS; LENGTH; PLASMA; PLASMA DENSITY; PLASMA DIAGNOSTICS; VISIBLE RADIATION
- Descriptors DEC
- DIMENSIONS; ELECTROMAGNETIC RADIATION; MEASURING INSTRUMENTS; RADIATIONS
Optional Information
- Notes
- (c) 2006 American Institute of Physics