Published August 2006 | Version v1
Journal article

Heterodyne interferometer with unequal path lengths

  • 1. Applied Physics, California Institute of Technology, Pasadena, California 91125 (United States)

Description

Laser interferometry is an extensively used diagnostic for plasma experiments. Existing plasma interferometers are designed on the presumption that the scene and reference beam path lengths have to be equal, a requirement that is costly in both the number of optical components and the alignment complexity. It is shown here that having equal path lengths is not necessary, instead, what is required is that the path length difference be an even multiple of the laser cavity length. This assertion has been verified in a heterodyne laser interferometer that measures typical line-average densities of ∼1021/m2 with an error of ∼1019/m2

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
77
Journal Issue
8
Journal Page Range
p. 083503-083503.6
ISSN
0034-6748
CODEN
RSINAK

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
38026496
Subject category
S70: PLASMA PHYSICS AND FUSION TECHNOLOGY; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
ALIGNMENT; ERRORS; INTERFEROMETERS; INTERFEROMETRY; LASER CAVITIES; LASERS; LENGTH; PLASMA; PLASMA DENSITY; PLASMA DIAGNOSTICS; VISIBLE RADIATION
Descriptors DEC
DIMENSIONS; ELECTROMAGNETIC RADIATION; MEASURING INSTRUMENTS; RADIATIONS

Optional Information

Notes
(c) 2006 American Institute of Physics