Published September 1999 | Version v1
Journal article

Artefact on MRA following aneurysm clipping: an in vitro study and prospective comparison with conventional angiography

  • 1. University Department of Neurosurgery, Institute of Neurology, London (United Kingdom)
  • 2. Lysholm Department of Neuroradiology, The National Hospital for Neurology and Neurosurgery, London (United Kingdom)

Description

Using both an experimental model and clinical cases, we looked at the artefact produced by Aesculap titanium-alloy aneurysm clips on MRA. Experimentally, the volume affected by artefact was 50 % less when the clip was imaged lying parallel to the main ferromagnetic field than when lying perpendicular to it. Clinically, MRA was prospectively compared with digital subtraction angiography (DSA) in nine patients who had undergone aneurysm clipping. One patient with a non-diagnostic MRA due to movement artefact was excluded. In all other cases there was an area of signal loss surrounding the clips, obscuring the immediately adjacent vessel segments. There was good demonstration of the adjacent bifurcations in five cases and the contralateral circulation was seen well in all patients. In three cases in which the adjacent bifurcations were not seen, considerable vasospasm was suggested by MRA and confirmed with DSA. In one patient an unclipped contralateral ophthalmic artery aneurysm was identified using both modalities. In this series there were no adverse events relating to clips in either static or time-varying magnetic fields. (orig.)

Additional details

Publishing Information

Journal Title
Neuroradiology
Journal Volume
41
Journal Issue
9
Journal Page Range
p. 680-686
ISSN
0028-3940
CODEN
NRDYAB

INIS

Country of Publication
Germany
Country of Input or Organization
Germany
INIS RN
30051125
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BIOMEDICAL RADIOGRAPHY; BLOOD VESSELS; COMPARATIVE EVALUATIONS; MAGNETIC FIELDS; NMR IMAGING
Descriptors DEC
BODY; CARDIOVASCULAR SYSTEM; DIAGNOSTIC TECHNIQUES; EVALUATION; MEDICINE; ORGANS

Optional Information

Notes
With 5 figs., 1 tab., 24 refs.