Comparison of central corneal thickness measurements between anterior chamber-optical coherence tomography and ultrasonic pachymeter

  • 1. Department of Ophthalmology, First Hospital, Jilin University, Changchun (China)

Description

Objective: To assess the agreement of central corneal thickness (CCT) measured by anterior chamber-optical coherence tomography (AC-OCT) and ultrasonic pachymeter and provide an objective basis for clinical application of AC-OCT. Methods: CCT of 150 college student volunteers (300 eyes) measured by two devices were obtained. The data was analyzed by paired t test and Pearson correlation analysis. Bland-Altman plot and Mountain plot were used to assess the agreement. Results: The mean CCT values were (530.05 ± 33.611) μm measured by AC-OCT and (543.68 ± 35.088) μm measured by ultrasonic pachymeter. Regression analysis showed a high correlation between the values obtained by both devices (r=0.960, P<0.001). Compared with AC-OCT, ultrasonic pachymeter overestimated the CCT by a mean of 13.62 μm. The two modalities had incomparable results. Conclusion: It is important to be noted in clinical practice that the measurements acquired by these two modalities are not directly interchangeable. However, the CCT measurements by the AC-OCT and ultrasonic pachymeter are highly correlated. AC-OCT is an effective method to observe the changes of the corneal thickness in the long term. (authors)

Additional details

Publishing Information

Journal Title
Journal of Jilin University. Medicine Edition
Journal Volume
36
Journal Issue
4
Journal Page Range
p. 775-778
ISSN
1671-587X

Optional Information

Notes
3 figs., 1 tab., 11 refs.