Published July 15, 1983 | Version v1
Journal article

Progress in energy resolution of mercuric iodide (HgI2) X-ray spectrometers

  • 1. University of Southern California, Marina Del Rey (USA). Inst. for Physics and Imaging Science
  • 2. Tracor Xray Inc. Mountain View, CA (USA)

Description

A brief discussion of the main factors which affect the energy resolution of HgI2 X-ray spectrometers is presented. A short review of the progress made in the process of improving the energy resolution is given. The latest experimental results obtained with a state-of-the-art cryogenically cooled low-noise preamplifier are shown. Ways of lowering the currently attainable room-temperature electronic noise level other than cryogenic cooling are discussed. (orig.)

Additional details

Publishing Information

Journal Title
Nucl. Instrum. Methods Phys. Res.
Journal Volume
213
Journal Issue
1
Series
Nucl. Instrum. Methods Phys. Res.
Journal Page Range
89-94
ISSN
0029-554X

Conference

Title
5. international workshop on mercuric iodide nuclear radiation detectors.
Dates
6-8 Jun 1982.
Place
Jerusalem (Israel).

Optional Information

Contract/Grant/Project number
Contract NSG-7535; DE-AM03-76SF00113