Published March 2000 | Version v1
Journal article

Determination of absorbed dose to water in reference conditions for radiotherapy kilovoltage x-rays between 10 and 300 kV: a comparison of the data in the IAEA, IPEMB, DIN and NCS dosimetry protocols

  • 1. Dosimetry and Medical Radiation Physics Section, International Atomic Energy Agency, Wagramer Strasse 5, A-1400 Vienna (Austria)

Description

A comparison of four of the most commonly used dosimetry protocols for the determination of absorbed dose to water in therapeutic kilovoltage x-rays using an ionization chamber (IAEA TRS-277, IPEMB, DIN and NCS) has been carried out. Owing to the different energy ranges and HVLs recommended by each protocol, backscatter factors, water-to-air mass energy absorption coefficient ratios and perturbation correction factors have been recast to a common quality range that all protocols satisfy individually to make a comparison possible. The results of the comparison show that in the sometimes reduced quality range originally included by the different protocols, determinations of absorbed dose to water at all beam qualities agree to within ±1.0% with that obtained using the second edition of the IAEA TRS-277 code of practice (1997). The extrapolation of data to a common beam quality range practically preserves the agreement for all the protocols except for that issued by the NCS at the extremes of the range, where differences of up to 1.8% and 1.4% have been found for low and medium energies respectively. In all cases the DIN protocol yields very good agreement with TRS-277. (author)

Additional details

Publishing Information

Journal Title
Physics in Medicine and Biology (Online)
Journal Volume
45
Journal Issue
3
Journal Page Range
p. 563-575
ISSN
1361-6560

INIS

Country of Publication
International Atomic Energy Agency (IAEA)
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
31022244
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
COMPARATIVE EVALUATIONS; DOSIMETRY; ENERGY RANGE; IONIZATION CHAMBERS; RADIATION DOSES; RADIOTHERAPY; RECOMMENDATIONS; WATER; X RADIATION
Descriptors DEC
ELECTROMAGNETIC RADIATION; EVALUATION; HYDROGEN COMPOUNDS; IONIZING RADIATIONS; MEASURING INSTRUMENTS; MEDICINE; OXYGEN COMPOUNDS; RADIATION DETECTORS; RADIATIONS; THERAPY