Precision of image-guided radiotherapy (IGRT) in six degrees of freedom and limitations in clinical practice

  • 1. Wuerzburg Univ. (Germany). Dept. of Radiation Oncology

Description

Purpose: To evaluate the precision of image-guided radiotherapy (IGRT) using cone-beam computed tomography (CB-CT) for volume imaging and a robotic couch for correcting setup errors in six degrees of freedom. Patients and Methods: 47 consecutive patients with 372 fractions were classified according to whether a patient fixation device was used (patfix: n = 28) or not (patnon-fix: n = 19). Prior to treatment a CB-CT was acquired and translational and rotational setup errors were corrected online without an action level using a robotic couch (HexaPOD). A second CB-CT was acquired after the correction process and after treatment in 134 and 238 fractions, respectively. Results: In 17 fractions (4.6%) rotational errors > 3 exceeded the motion range of the HexaPOD. Errors (3D vector) after the correction process were significantly smaller for patfix compared to patnon-fix (p < 0.001): 0.9 mm ± 0.5 mm and 1.6 mm ± 0.8 mm, respectively. For patnon-fix the correction of rotational errors resulted in displacements of the patients on the angled couch of 0.6 mm/1 . Intrafractional motion further decreased precision in patnon-fix but not in patfix. Conclusion: Very high precision in cranial and extracranial treatment of immobilized patients was demonstrated. Without application of adequate immobilization the correction of rotational errors and intrafractional patient motion significantly decreased the accuracy of the online correction protocol. (orig.)

Additional details

Publishing Information

Journal Title
Strahlentherapie und Onkologie
Journal Volume
183
Journal Issue
6
Journal Page Range
p. 307-313
ISSN
0179-7158
CODEN
STONE4