Wide-range neutron dose determination with CR-39
Creators
- 1. Atomic Energy of Canada Ltd., Chalk River, ON (Canada). Chalk River Nuclear Labs.
Description
Optical density measurements of CR-30 irradiated with 252Cf neutrons and chemically etched with 6.5 N KOH solution have been used to determine neutron absorbed doses between 0.1 and 10 Gy. Optimum etching conditions will depend upon the absorbed dose. Since it is not always possible to know the range of absorbed dose on a CR-39 dosemeter collected from personnel and area monitor stations in a criticality accident situation, a three-step two-hour chemical etch at 60oC has been found to be appropriate. If after a total of six hours of chemical etching the optical density is found to be below 0.04 for 500 nm light (transmission > 90%) then further treatment in the form of electrochemical etching can be carried out to determine the lower absorbed dose. In this manner, absorbed doses below 0.1 Gy can be determined by counting tracks over a unit area. (author)
Additional details
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 58
- Journal Issue
- 3
- Journal Page Range
- p. 201-204.
- ISSN
- 0144-8420
- CODEN
- RPDODE
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 26041700
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- DIELECTRIC TRACK DETECTORS; ETCHING; NEUTRON DOSIMETRY; OPACITY
- Descriptors DEC
- DOSIMETRY; MEASURING INSTRUMENTS; OPTICAL PROPERTIES; PHYSICAL PROPERTIES; RADIATION DETECTORS