Published 2007 | Version v1
Journal article

Considerations concerning the influence of magnetic field on calibration of thermometers for low and very low temperature measurements

  • 1. University of Craiova, 61 Unirii Street, Craiova (Romania)

Description

Based on the influence of the magnetic fields on the resistivity of the thermometers used in low and very low temperature measurements some aspects of the traceability of the calibration of these thermometers are investigated. Results are given for resistance thermometers of platinum, carbon, thermistors, germanium, silicon and gallium arsenide for the temperature ranges (13K-273K), (0.02K- 100K), (4K-273K), (0.05-100K), (1K-400K) and (1K-273K), respectively as well as for thermoelectric thermometers with copper-constantan thermocouple and gold-iron (chromium) thermocouple for the temperature ranges (10K-273K) and (91K-273K), respectively

Availability note (English)

Available from author(s) or National Institute of Research-Development for Cryogenics and Isotopic Technologies, ICSI, PO 4, PO Box 10, 4 Uzinei Str, RO-240050 Rm. Valcea (RO)); see also

Additional details

Publishing Information

Journal Title
Progress of Cryogenics and Isotopes Separation
Journal Volume
10
Journal Issue
19-20
Journal Page Range
p. 17-22
ISSN
1582-2575

Optional Information

Notes
4 refs., 3 figs., 1 tab.