Published August 2013 | Version v1
Journal article

Research on dead layer thickness and cold finger size of coaxial HPGe detector

  • 1. College of Physics and Electronic Information, Inner Mongolia University for Nationalities, Tongliao (China)

Description

HPGe detector crystal dead layer thickness and cold finger size were corrected by Monte Carlo simulation along with experiment. The results show that when the HPGe detector dead layer thickness is 0.22 cm, the thickness and length of cold-finger size are 0.301 and 1 cm, the simulated efficiency is consistent with the experimental efficiency. (authors)

Additional details

Publishing Information

Journal Title
Nuclear Electronics and Detection Technology
Journal Volume
33
Journal Issue
8
Journal Page Range
p. 960-963
ISSN
0258-0934

Optional Information

Notes
7 figs., 3 refs.