Published 1995 | Version v1
Report Open

Source imaging of drums in the APNEA system

Description

The APNea System is a neutron assay device utilizing both a passive mode and a differential-dieaway active mode. The total detection efficiency is not spatially uniform, even for an empty chamber, and a drum matrix in the chamber can severely distort this response. In order to achieve a response which is independent of the way the source material is distributed in a drum, an imaging procedure has been developed which treats the drum as a number of virtual (sub)volumes. Since each virtual volume of source material is weighted with the appropriate instrument parameters (detection efficiency and thermal flux), the final assay result is essentially independent of the actual distribution of the source material throughout the drum and its matrix

Availability note (English)

MF available from INIS under the Report Number; Also available from OSTI as DE96005981; NTIS; US Govt. Printing Office Dep.

Files

27048437.pdf

Files (526.0 kB)

Name Size Download all
md5:39d38b1fd8a2db7d4bf2954dbe2b4a44
526.0 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
22 p.
Report number
CONF-951091--10

Conference

Title
4. nondestructive assay and nondestructive examination waste characterization conference.
Dates
24-26 Oct 1995.
Place
Salt Lake City, UT (United States).

Optional Information

Contract/Grant/Project number
Contract AC05-84OR21400
Funding organization
USDOE, Washington, DC (United States).