Combined proton-recoil and neutron time-of-flight spectrometer for 14 MeV neutrons

Description

The main effort put into this work is the foundation of a reliable physical basis for a 12-16 MeV neutron-spectrometer at JET. The essential problem is the amount of scatterer that can be incorporated without losing resolution. We have found two possible methods, the use of a pure hydrogen scatterer and the use of a polyethylene foil scatterer. The pure hydrogen solution gives a very complicated spectrometer with large detectors. The polyethylene solution is limited by the thickness and the width of the foil. We judge the solution with the polyethylene foil to be the most promising one for a reliable spectrometer. However, a large foil area is needed. This gives a spectrometer design with an annular foil, an annular neutron detection system, and a central proton-detector. An efficiency of 10-6 counts/s per n/cm2,s at the foil can be obtained with a resolution in the order of 100 keV for 14 MeV neutrons. Following the General Requirements given in the contract of this work, we concluded that an instrument with the desired properties can be made. The instruments is able to give useful information about the plasma from plasma temperatures of about 5 keV. (Authors)

Availability note (English)

MF available from INIS under the Report Number.

Files

14797424.pdf

Files (2.9 MB)

Name Size Download all
md5:720a1da7457f2bf3426cf6e16f230328
2.9 MB Preview Download

Additional details

Publishing Information

Imprint Pagination
163 p.
Report number
CTH-RF--43

Optional Information

Contract/Grant/Project number
Contract JET JB2/9008