Published 2011 | Version v1
Miscellaneous Open

Uncertainty Measurement of kVp Output of the X-ray System using Spectrometry System

  • 1. Malaysian Nuclear Agency, Bangi, Kajang, Selangor (Malaysia), Radiation Safety and Health Div.

Description

This study was carried out to determine the uncertainty of kVp output of the x-ray tube used in the calibration of non-invasive kVp meters. A non-invasive method was used to measure the spectrum of the x-ray output using Amptek XR-100T-CdTe spectrometry system. While an invasive high voltage divider (dynalyser) coupled to the x-ray system measures the true kilo voltage supplied to the x-ray tube with uncertainty of 2.5 % (k=2). The consistency of the kVp output was monitored daily at 9 points ranging between 40 kV-120 kV with interval steps of 10 kV from the dynalyser system. While the x-ray output spectrum of the 9 points were measured once a year. The uncertainty was determined from the consistency x-ray output, dynalyser system uncertainty, the spectrometry system accuracy, error and variation. The test results showed that kVp output measured by the dynalyser system everyday is consistent with coefficient of variations of not more than 0.89 %. The kVp output via dynalyser system and Spectrometry system shows good agreement with standard error of not more than 2.48 %. The total uncertainty of kVp output of the x-ray system using the spectrometry system is not more than 6.2 % (k=2). As a conclusion, the factors influencing the quality and quantity of the x-ray output also influence the uncertainty of the kVp output. (author)

Availability note (English)

Also available in Malaysian Nuclear Agency Document Delivery Center

Files

44122688.pdf

Files (234.6 kB)

Name Size Download all
md5:ac859de793534ca0c423556810c0c259
234.6 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
3 p.
Report number
INIS-MY--2013-170

Conference

Title
Nuclear Technical Convention 2011
Acronym
NTC 2011
Dates
13-15 Sep 2011
Place
Bangi (Malaysia)

INIS

Country of Publication
Malaysia
Country of Input or Organization
Malaysia
INIS RN
44122688
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ELECTRIC POTENTIAL; MALAYSIA; MINT; SPECTROMETERS; UNCERTAINTY PRINCIPLE; X-RAY EQUIPMENT
Descriptors DEC
ASIA; DEVELOPING COUNTRIES; EQUIPMENT; MALAYSIAN ORGANIZATIONS; MEASURING INSTRUMENTS; NATIONAL ORGANIZATIONS

Optional Information

Notes
Oral presentation. 16 slides