Description

A determination of the exact statistical limits in the accuracy of time measurements in scintillation counters is presented. An expression for the variance of the timing distribution, where the number of photons in the signal is fixed, is first derived. This leads to an expression for the case where the total number of photons in the emission is randomized in accordance with Poisson statistics

Additional details

Identifiers

DOI
10.1016/j.nima.2006.04.049;
PII
S0168-9002(06)00693-0;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
564
Journal Issue
1
Journal Page Range
p. 185-189
ISSN
0168-9002
CODEN
NIMAER

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
38035068
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
ACCURACY; DISTRIBUTION; PHOTON EMISSION; PHOTONS; SCINTILLATION COUNTERS; SIGNALS; STATISTICS; TIME MEASUREMENT
Descriptors DEC
BOSONS; ELEMENTARY PARTICLES; EMISSION; MASSLESS PARTICLES; MATHEMATICS; MEASURING INSTRUMENTS; RADIATION DETECTORS

Optional Information

Copyright
Copyright (c) 2006 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.