Published 1987 | Version v1
Report Open

Method for continuous registration of linear spectrometric circuit alive time

Description

A method for continuous registration of the alive time of linear spectrometric circuit in regime of pulse counting is developed for measuring gamma spectra in neutron resonances. The neutron energy in these measurements is determined by the time-of-flight method. The alive time of the circuit is registrated by means of pulses of a periodic generator with stable amplitude which receive an additional status bit in ADC. The information on amplitude of the detector pulse and on time of flight of each event is recorded on the magnetic tape. The time of flight spectrum of generator pulses is obtained when scanning information by additional status bit. The analysis of this spectrum permits to introduce corrections on the alive time of spectrometric circuit

Files

19033288.pdf

Files (130.7 kB)

Name Size Download all
md5:c92ec09a4f7303409a49bbf15d54a980
130.7 kB Preview Download

System files (17.7 kB)

Name Size Download all

Additional details

Additional titles

Original title (Russian)
Методика непрерывной регистрации живого времени линейного спектрометрического тракта

Publishing Information

Imprint Pagination
4 p.
Report number
JINR-R--13-87-144

Optional Information

Notes
3 refs.; 2 figs.