Published April 1987 | Version v1
Report Restricted

Calculated response of a 5.5 x 5.5 cm high-purity Ge detector to gamma rays with energies up to 20 MeV

Description

Monte Carlo calculations have been done of the energy and angular response of a 5.5 x 5.5 cm, high-purity Ge detector, for gamma rays incident with energies from 0.1 to 20 MeV. Results are given for the absolute probabilities of total energy absorption and of single and double annihilation escape, which relate the areas of the peaks in the measured pulse-height distribution to the intensities of incident gamma-ray lines

Availability note (English)

Available from NTIS, PC A02/MF A01; 1 as DE87013947.

Files

Restricted

The record is publicly accessible, but files are restricted to users with access.

Additional details

Publishing Information

Imprint Pagination
12 p.
Report number
NBSIR--87-3548

Optional Information

Notes
Portions of this document are illegible in microfiche products.