Published November 2018 | Version v1
Journal article

Spectroscopic Measurements of L X-rays with a TES Microcalorimeter for a Non-destructive Assay of Transuranium Elements

  • 1. Japan Atomic Energy Agency (Japan)
  • 2. Kyushu University (Japan)
  • 3. Japan Aerospace Exploration Agency (Japan)

Description

Spectroscopic measurement of the L X-rays emitted from transuranium elements is one of the most useful techniques for the non-destructive assays of nuclear materials. In this study, we fabricated a transition-edge-sensor (TES) microcalorimeter using a 5-μm-thick Au absorber and tested its ability to measure the L X-rays emitted from two transuranium elements, Np-237 and Cm-244 sources. The microcalorimeter was found to successfully measure the L X-rays with an energy resolution (full width at half maximum) below 70 eV. These results confirm that L X-rays can be identified using the proposed TES microcalorimeter to enable non-destructive assays of transuranium elements.

Additional details

Identifiers

Publishing Information

Journal Title
Journal of Low Temperature Physics
Journal Volume
193
Journal Issue
3-4
Journal Page Range
p. 314-320
ISSN
0022-2291
CODEN
JLTPAC

Conference

Title
International workshop on low temperature detectors
Acronym
LTD17
Dates
17-21 Jul 2017
Place
Kurume City, Fukuoka (Japan)

Optional Information

Copyright
Copyright (c) 2018 Springer Science+Business Media, LLC, part of Springer Nature
Notes
http://www.springer-ny.com