Published March 2012 | Version v1
Journal article

Control chart analysis of data from a multicenter monitor unit verification study

  • 1. Department of Medical Radiation Physics, Lund University, Malmö (Sweden)
  • 2. Department of Radiation Physics, Skåne University Hospital, Malmö (Sweden)
  • 3. Department of Radiation Physics, Kalmar County Hospital (Sweden)
  • 4. Department of Medical Radiation Physics, Lund University, Lund (Sweden)

Description

Background and purpose: This study aims to investigate the process of monitor unit verification using control charts. Control charts is a key tool within statistical process control (SPC), through which process characteristics can be visualized, usually chronologically with statistically determined limits. Material and methods: Our group has developed a monitor unit verification software that has been adopted at several Swedish institutions for pre-treatment verification of radiotherapy treatments. Deviations between point dose calculations using the treatment planning systems and using the independent monitor unit verification software from 9219 treatment plans and five different institutions were included in this multicenter study. The process of monitor unit verification was divided into subprocesses. Each subprocess was analyzed using probability plots and control charts. Results: Differences in control chart parameters for the investigated subprocesses were found between different treatment sites and different institutions, as well as between different treatment techniques. 19 of 37 subprocesses met the clinical specification (±5%), i.e. process capability index was equal to or above one. Conclusions: Control charts were found to be a useful tool for continuous analysis of data from the monitor unit verification software for patient specific quality control, as well as for comparisons between different institutions and treatment sites. The derived control chart limits were in agreement with AAPM TG114 guidelines on action levels.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.radonc.2011.11.016

Additional details

Identifiers

DOI
10.1016/j.radonc.2011.11.016;
PII
S0167-8140(11)00715-8;

Publishing Information

Journal Title
Radiotherapy and Oncology
Journal Volume
102
Journal Issue
3
Journal Page Range
p. 364-370
ISSN
0167-8140
CODEN
RAONDT

INIS

Country of Publication
Ireland
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
44102042
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
COMPUTER CODES; DATA ANALYSIS; PATIENTS; PLANNING; PROBABILITY; PROCESS CONTROL; QUALITY CONTROL; RADIATION DOSES; RADIOTHERAPY; RECOMMENDATIONS; SPECIFICATIONS; VERIFICATION
Descriptors DEC
CONTROL; DOSES; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; THERAPY

Optional Information

Copyright
Copyright (c) 2011 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.