Published June 1975 | Version v1
Journal article

Error introduced in the radiocardiographic measurement of right volume by the non-linearity of the detector in high count rates

  • 1. Centre Hospitalier Regional Augustin-Morvan, 29N - Brest (France)

Description

The short life tracers permit the use of high activities and so allow to follow with a good statistic accuracy fast phenomena such as the different phases of cardiac revolution. However, this advantage may be quickly limited by the dead-time of the detection unit used. It results from this, in high activities, a non-linear response which introduces non negligible errors which have been calculated and experimentally verified in the measurement of right ventricular volumes

Abstract (French)

Les traceurs radioactifs a vie breve autorisent l'usage d'activites elevees et permettent ainsi de suivre, avec une bonne precision statistique, des phenomenes rapides tels que les differentes phases de la revolution cardiaque. Cependant, cet avantage peut etre rapidement limite par le temps mort de la chaine de detection utilisee. Il en resulte, aux fortes activites, une reponse non lineaire qui introduit des erreurs non negligeables qui ont ete calculees et verifiees experimentalement dans le cas de la mesure des volumes ventriculaires droits

Additional details

Additional titles

Original title (French)
Erreur introduite sur la mesure radiocardiographique du volume cavitaire droit par la non-linearite du detecteur aux forts taux de comptage

Publishing Information

Journal Title
J. Radiol., Electrol., Med. Nucl.
Journal Volume
56
Journal Issue
6
Series
J. Radiol., Electrol., Med. Nucl.
Journal Page Range
517-522

INIS

Country of Publication
France
Country of Input or Organization
France
INIS RN
7240258
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
DEAD TIME; ERRORS; RADIOCARDIOGRAPHY; RADIOISOTOPE SCANNERS
Descriptors DEC
CARDIOGRAPHY; DIAGNOSTIC TECHNIQUES; TIMING PROPERTIES