Published September 1987
| Version v1
Journal article
Stabilization of the photomultiplier gain of a liquid scintillation counter
Description
A stabilization system of photomultiplier gain, where light-emitting diode flashes have been used to obtain a reference signal, is described. The diode is placed just in the liquid scintilllator volume. The stabilization system contains several (according to the number of photomultipliers) identical channels, which of them consists of a colorimeter, a control trigger and an integrator with an operational amplifier. Increase of photomultiplier stability is reached by changing voltage of photomultiplier power according to the reference signal amplitude. The level of background and efficiency of neutron detection by a scintillation counter are unchanged when using the stabilization system for 10 days of measurements
Additional details
Additional titles
- Original title (Russian)
- Стабилизация усиления фотоумножителей жидкостного сцинтилляционного счетчика
Publishing Information
- Journal Title
- Prib. Tekh. Ehksp.
- Journal Issue
- no.5
- Series
- Prib. Tekh. Ehksp.
- ISSN
- 0032-8162
- CODEN
- PRTEA
INIS
- Country of Publication
- Russian Federation
- Country of Input or Organization
- USSR
- INIS RN
- 19069057
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- FLOWSHEETS; GAIN; LIGHT EMITTING DIODES; LIQUID SCINTILLATION DETECTORS; PHOTOMULTIPLIERS; STABILITY
- Descriptors DEC
- AMPLIFICATION; DIAGRAMS; INFORMATION; MEASURING INSTRUMENTS; PHOTOTUBES; RADIATION DETECTORS; SCINTILLATION COUNTERS; SEMICONDUCTOR DEVICES; SEMICONDUCTOR DIODES