Published October 1, 2003 | Version v1
Journal article

X-ray stereotactic lesion localization in conjunction with dedicated scintimammography

Description

We are developing a dual modality system that combines digital X-ray mammography with gamma emission scintigraphy on an upright mammography gantry. The breast is held under mild compression by a support structure that is independent of the detectors. The X-ray source and detectors can be rotated around a fixed rotation axis permitting multiple views of the breast with fixed compression. Two such views can be combined as a stereotactic pair to obtain the three-dimensional location of breast lesions. Information about the location of the lesion within the breast permits corrections for attenuation and detector spatial resolution, resulting in more accurate estimation of the true lesion-to-background concentration ratio, based on the image lesion-to-background counts ratio. In this paper, we describe the model used to make these corrections, and present the results of the phantom experiments designed to test the accuracy of our calculations

Additional details

Publishing Information

Journal Title
IEEE Transactions on Nuclear Science
Journal Volume
50
Journal Issue
5
Journal Page Range
p. 7
ISSN
0018-9499
CODEN
IETNAE

Optional Information

Contract/Grant/Project number
AC--05-84ER40150
Notes
IEEE Transactions on Nuclear Science, Volume 50, Issue 5, Page(s) 1636 - 1642
Funding organization
USDOE Office of Energy Research ER (United States)
Secondary number(s)
JLAB-PHY--03-59; DOE/ER--40150-3004