NaI(Tl) response functions
Creators
- 1. Universidad Autonoma de Zacatecas, Unidad Academica de Estudios Nucleares, Cipres No. 10, Fracc. La Penuela, 98068 Zacatecas, Zac. (Mexico)
- 2. Centro Estatal de Cancerologia de Nayarit, Calz. de la Cruz 118 Sur, Tepic, Nayarit (Mexico)
- 3. Instituto Tecnologico de Aguascalientes, Av. Adolfo Lopez Mateos 1801 Ote., 20155 Aguascalientes, Ags. (Mexico)
Description
The response functions of a NaI(Tl) detector have been estimated using Monte Carlo methods. Response functions were calculated for monoenergetic photon sources (0.05 to 3 MeV). Responses were calculated for point-like sources and for sources distributed in Portland cement cylinders. The responses were used to calculate the efficiency functions in term of photon energy. Commonly, sources used for calibration are point-like, and eventually sources to be measured have different features. In order to use the calibrated sources corrections due to solid angle, self-absorption and scattering, must be carried out. However, some of these corrections are not easy to perform. In this work, the calculated responses were used to estimate the detector efficiency of point-like sources, and sources distributed in Portland type cement. Samples of Portland paste were prepared and were exposed to photoneutrons produced by a 15 MV linac. Some of the elements in the cement were activated producing γ-emitting radionuclides that were measured with a NaI(Tl) gamma-ray spectrometer, that was calibrated with point-like sources. In order to determine the specific activity in the induced radioisotopes calculated efficiencies were used to make corrections due to the differences between the solid angle, photon absorption and photon scattering in the point-like calibration sources and the sources distributed in cement. During the interaction between photoneutrons and the cement samples three radioisotopes were induced: 56Mn, 24Na, and 28Al. (Author)
Availability note (English)
Available from Instituto Nacional de Investigaciones Nucleares, Centro de Informacion y Documentacion, 52750 Ocoyoacac, Estado de Mexico (MX), e-mail: mclaudia.gonzalez@inin.gob.mxAdditional details
Publishing Information
- Publisher
- Sociedad Nuclear Mexicana
- Imprint Place
- Ciudad de Mexico (Mexico)
- Imprint Pagination
- 10 p.
Conference
- Title
- nuclear energy, climate change, and safety challenges and radiological protection
- Original Conference Title
- 26. Congreso Anual de la SNM
- Acronym
- 26. SNM Annual Congress; 14. SMSR National Congress
- Dates
- 5-8 Jul 2015
- Place
- Puerto Vallarta, Jalisco (Mexico)
INIS
- Country of Publication
- Mexico
- Country of Input or Organization
- Mexico
- INIS RN
- 47069212
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference, Non-conventional Literature
- Descriptors DEI
- ALUMINIUM 28; CALIBRATION; CORRECTIONS; CYLINDERS; EFFICIENCY; GAMMA SPECTROMETERS; INTERACTIONS; LINEAR ACCELERATORS; MANGANESE 56; MEV RANGE; MONTE CARLO METHOD; PHOTONEUTRONS; PHOTONS; PORTLAND CEMENT; RESPONSE FUNCTIONS; SCATTERING; SELF-ABSORPTION; SODIUM 24; SODIUM IODIDES
- Descriptors DEC
- ABSORPTION; ACCELERATORS; ALKALI METAL COMPOUNDS; ALUMINIUM ISOTOPES; BARYONS; BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; BOSONS; BUILDING MATERIALS; CALCULATION METHODS; CEMENTS; ELEMENTARY PARTICLES; ENERGY RANGE; FERMIONS; FUNCTIONS; HADRONS; HALIDES; HALOGEN COMPOUNDS; HOURS LIVING RADIOISOTOPES; INORGANIC PHOSPHORS; INTERMEDIATE MASS NUCLEI; IODIDES; IODINE COMPOUNDS; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; LIGHT NUCLEI; MANGANESE ISOTOPES; MASSLESS PARTICLES; MATERIALS; MEASURING INSTRUMENTS; MILLISECONDS LIVING RADIOISOTOPES; MINUTES LIVING RADIOISOTOPES; NEUTRONS; NUCLEI; NUCLEONS; ODD-ODD NUCLEI; PHOSPHORS; PHOTONUCLEONS; RADIOISOTOPES; SODIUM COMPOUNDS; SODIUM HALIDES; SODIUM ISOTOPES; SORPTION; SPECTROMETERS