Published June 21, 2015 | Version v1
Journal article

Simultaneous EEG–fMRI: evaluating the effect of the cabling configuration on the gradient artefact

  • 1. Sir Peter Mansfield Magnetic Resonance Centre, School of Physics and Astronomy, University of Nottingham, Nottingham, NG2 5HU (United Kingdom)

Description

EEG recordings made in combined EEG–fMRI studies are corrupted by gradient artefacts (GAs) resulting from the interaction of the EEG system with the time-varying magnetic field gradients used in MRI. The dominant contribution to the GA arises from interaction with the leads of the EEG cap and the human head, but artefacts are also produced in the cables used to connect the EEG cap to the amplifier. The aim of this study is to measure the effects of the connecting cable configuration on the characteristics of the GA. We measured the GA produced on two different cable configurations (a ribbon cable and a cable consisting of wires that are twisted together to form a cylindrical bundle) by gradient pulses applied on three orthogonal axes and also characterized the effect of each cable configuration on the GA generated by a multi-slice echo planar imaging sequence, as employed in typical EEG–fMRI studies. The results demonstrate that the cabling that connects the EEG cap to the amplifier can make a significant contribution to the GA recorded during EEG–fMRI studies. In particular, we demonstrate that the GA generated by a ribbon cable is larger than that produced using a twisted cable arrangement and that changes in the GA resulting from variation in the cable position are also greater for the ribbon cable. (note)

Availability note (English)

Available from http://dx.doi.org/10.1088/0031-9155/60/12/N241

Additional details

Identifiers

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
60
Journal Issue
12
Journal Page Range
p. N241-N250
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
47072521
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
AMPLIFIERS; BIOMEDICAL RADIOGRAPHY; HEAD; IMAGE PROCESSING; MAGNETIC FIELDS; NMR IMAGING
Descriptors DEC
BODY; DIAGNOSTIC TECHNIQUES; ELECTRONIC EQUIPMENT; EQUIPMENT; MEDICINE; NUCLEAR MEDICINE; PROCESSING; RADIOLOGY