An investigation of the possible application of a bidimensional MWPC to x-ray absorptiometry
Description
The advantages and disadvantages of using a xenon filled MWPC X-ray detector with a 153Gd X-ray source are examined relative to the standard scanning method using a NaI(Tl) crystal. The questions of statistical precision, spatial resolution and energy resolution are examined experimentally and theoretically. While the position sensitivity of the device offers assistance with the difficult problem of repositioning, its poor energy resolution leads to rapid saturation of the measured absorption. It is suggested that this problem can be solved by the use of filters and/or collimation on the MWPC. Given the solution of this problem it appears that the technique can yield an accuracy in the 1% region for bone mineral measurements. (author)
Availability note (English)
MF available from INIS under the Report Number.
Files
Additional details
Publishing Information
- Imprint Pagination
- 14 p.
- Report number
- RL--76-067
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 8281837
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ABSORPTION; BIOMEDICAL RADIOGRAPHY; CHEMICAL ANALYSIS; GADOLINIUM 153; MINERALS; MULTIWIRE PROPORTIONAL CHAMBER; RESOLUTION; SENSITIVITY; SKELETON; X RADIATION; X-RAY DETECTION; X-RAY SOURCES
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BODY; DAYS LIVING RADIOISOTOPES; DIAGNOSTIC TECHNIQUES; ELECTROMAGNETIC RADIATION; ELECTRON CAPTURE RADIOISOTOPES; EVEN-ODD NUCLEI; GADOLINIUM ISOTOPES; INTERMEDIATE MASS NUCLEI; IONIZING RADIATIONS; ISOTOPES; MEASURING INSTRUMENTS; MEDICINE; NUCLEI; ORGANS; PROPORTIONAL COUNTERS; RADIATION DETECTION; RADIATION DETECTORS; RADIATION SOURCES; RADIATIONS; RADIOISOTOPES; RARE EARTH NUCLEI; X-RAY EQUIPMENT