Published 1988
| Version v1
Report
Proportional chambers with high uniform gas amplification for transition radiation detector
- 1. AN SSSR, Leningrad (USSR). Inst. Yadernoj Fiziki
Description
Measurement results of gas amplifications (GA) of proportional chambers being part of transition radiation detector are presented. The proportional chamber had 100x600 mm sensitive area with 8 mm anode-cathode gap. Anode wires were made of tungsten 25 mum in-diameter and were tightened with 2 mm spacing. Cathode planes were made of 4 mum thick aluminized mailar. The chamber was filled with Xe + 30% CH4 mixture. Eleven structural factors influencing GA fluctuation value that is 10% were considered. 2 refs.; 6 figs
Additional details
Additional titles
- Original title (Russian)
- Пропорциональные камеры с высокой однородностью коэффициента газового усиления для детектора переходного излучения
Publishing Information
- Imprint Title
- International symposium on position detectors in high energy physics. Proceedings
- Imprint Pagination
- 404 p.
- Journal Page Range
- p. 189-193.
- Report number
- JINR-D--1-13-88-172
Conference
- Title
- International symposium on position detectors in high energy physics.
- Dates
- 22-25 Sep 1987.
- Place
- Dubna (USSR).
INIS
- Country of Publication
- USSR
- Country of Input or Organization
- USSR
- INIS RN
- 21003409
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BINARY MIXTURES; FLUCTUATIONS; GAIN; METHANE; MULTIWIRE PROPORTIONAL CHAMBER; PERFORMANCE TESTING; TRANSITION RADIATION DETECTORS; XENON
- Descriptors DEC
- ALKANES; AMPLIFICATION; DISPERSIONS; ELEMENTS; HYDROCARBONS; MEASURING INSTRUMENTS; MIXTURES; NONMETALS; ORGANIC COMPOUNDS; PROPORTIONAL COUNTERS; RADIATION DETECTORS; RARE GASES; TESTING; VARIATIONS
Optional Information
- Notes
- Imprint:Mezhdunarodnyj simpozium po koordinatnym detektoram v fizike vysokikh ehnergij. Trudy.