Published November 1980
| Version v1
Journal article
X-ray beam characteristics of the Varian Clinac 6-100 linear accelerator
Description
Beam characteristics and dosimetry measurements of the 6 MV x-ray beam from a Varian Associates Clinac 6-100 linear accelerator are presented. Percentage depth dose tables are given as a function of field size for square fields. Tissue-maximum ratios and scatter-maximum ratios are tabulated as a function of depth and field size. Investigation of surface dose and depth of maximum dose reveal a dependence on field size. Other beam parameters measured are presented including field flatness, symmetry, SSD dependence, and penumbra. Treatment planning parameters including field size output factors and wedge factors are discussed
Additional details
Publishing Information
- Journal Title
- Med. Phys.
- Journal Volume
- 7
- Journal Issue
- 6
- Series
- Med. Phys.
- Journal Page Range
- 716-722
- ISSN
- 0094-2405
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 13669159
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE; S63: RADIATION, THERMAL, AND OTHER ENVIRONMENTAL POLLUTANT EFFECTS ON LIVING ORGANISMS AND BIOLOGICAL MATERIALS;
- Descriptors DEI
- BEAM DYNAMICS; DEPTH DOSE DISTRIBUTIONS; LINEAR ACCELERATORS; PLANNING; RADIATION DOSES; RADIOTHERAPY; X RADIATION; X-RAY DOSIMETRY; X-RAY RADIOGRAPHY
- Descriptors DEC
- ACCELERATORS; DOSIMETRY; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MEDICINE; RADIATION DOSE DISTRIBUTIONS; RADIATIONS; SPATIAL DOSE DISTRIBUTIONS; THERAPY