Published 2016 | Version v1
Book

Efficiency factors for Phoswich based lung monitor using ICRP Voxel phantoms

  • 1. Radiological Safety Division, Indira Gandhi Centre for Atomic Research, Kalpakkam (India)

Description

The actinide contamination in lungs is measured either using array of HPGe detector or Phoswich based lung monitors. This paper discusses the results obtained during numerical calibration of Phoswich based lung counting system using ICRP VOXEL phantoms. The results are also compared with measured efficiency values obtained using LLNL phantom. The efficiency factors of 241Am present in the lungs for phoswich detector was simulated using ICRP male voxel phantom and compared with experimentally observed values using LLNL Phantom. The observed deviation is 12%. The efficiency of the same for female subjects was estimated using ICRP female voxel phantom for both supine and posterior geometries

Part of:
Proceedings of the international conference on radiological safety in workplace, nuclear facilities and environment: book of abstracts

Additional details

Publishing Information

Publisher
Indian Association for Radiation Protection
Imprint Place
Mumbai (India)
Imprint Title
Proceedings of the international conference on radiological safety in workplace, nuclear facilities and environment: book of abstracts
Imprint Pagination
287 p.
Journal Page Range
p. 160

Conference

Title
32. international conference on radiological safety in workplace, nuclear facilities and environment
Acronym
IARPIC-2016
Dates
22-25 Feb 2016
Place
Kalpakkam (India)

Optional Information

Notes
2 refs., 1 fig., 1 tab.