Validation of neutron activation analysis method in education and training center
- 1. Pusat Pendidikan dan Pelatihan (PUSDIKLAT) - BATAN, JI. Lebak Bulus Raya, Pasar Jumat, Jakarta 12440 (Indonesia)
Description
Validation of Neutron Activation Analysis has been carried out, including validation of equipment and method. Validation of equipment that affect the results of the analysis includes an analytical balance and gamma spectrometer. While validation of the method includes confirmation of identity, selectivity and specificity, detection limit, repeatability and reproducibility, accuracy and precision. Validation of balance give results of good accuracy and precision of 0.0005 mg difference and the maximum standard deviation of 0.46 %. Validation of gamma spectrometer give good result and is discussed in Validation of Gamma Spectrometer in Spectroscopy Laboratory (Sugino, etc). The confirmation of identity shows Al, V, Cr K, Na, Ca, Zn, Mg, Mn, Fe, Co, Cu were analyzed base on Al-28, V-52, Cr-51, K-42, K-24, Ca-49, Zn-65, Mg-27, Mn-56, Fe-59, Co-60 and Cu-66. Detection limits of Al, V, Ca, Mg, Mn, As, K, Na, Mg, Ce, Co, Cr, Fe, La, Sc, Zn in sediment samples are 240 ppm, 27 ppm, 4750 ppm, 2600 ppm, 21 ppm, 3.3 ppm, 75 ppm, 1.4 ppm, 1.8 ppm, 0.5 ppm, 2.7 ppm, 29 ppm, 1 ppm, 0.05 ppm, 37 ppm. Detection limit of Cu is more than 98.6 ppm. Repeatability and reproducibility of sediment sample for Al, Cr, V, Mg, Zn, Co and As were good enough, but for Mn, Ca and Fe are relatively large differences. Result of Comparative Test for sediment samples indicate the accuracy and precision for Al, Co, Cr, Mn, V, As, Ca, Fe, Mg, Zn accepted. (author)
Files
48059291.pdf
Files
(2.0 MB)
| Name | Size | Download all |
|---|---|---|
|
md5:4b720a253cf1518c9db8644878edd6da
|
2.0 MB | Preview Download |
Additional details
Additional titles
- Original title (Indonesian)
- Validasi metode dalam analisis aktivasi neutron di Pusdiklat BATAN
Publishing Information
- Publisher
- National Nuclear Energy Agency
- Imprint Place
- Serpong (Indonesia)
- Imprint Title
- Proceedings of national seminar neutron activation analysis
- Imprint Pagination
- 351 p.
- Journal Page Range
- p. 125-130
- ISSN
- 2085-2797
- Report number
- INIS-ID--71
Conference
- Title
- National seminar neutron activation analysis
- Original Conference Title
- Seminar nasional analisis aktivasi neutron
- Dates
- 2-3 Nov 2010
- Place
- Serpong (Indonesia)
INIS
- Country of Publication
- Indonesia
- Country of Input or Organization
- Indonesia
- INIS RN
- 48059291
- Subject category
- S38: RADIATION CHEMISTRY, RADIOCHEMISTRY AND NUCLEAR CHEMISTRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ACCURACY; CALIBRATION STANDARDS; COMPARATIVE EVALUATIONS; ELEMENTS; GAMMA SPECTROSCOPY; LABORATORIES; NEUTRON ACTIVATION ANALYSIS; SENSITIVITY; SPECTROSCOPY; VALIDATION
- Descriptors DEC
- ACTIVATION ANALYSIS; CHEMICAL ANALYSIS; EVALUATION; NONDESTRUCTIVE ANALYSIS; SPECTROSCOPY; STANDARDS; TESTING
Optional Information
- Notes
- 9 refs.; 3 tabs.; 3 figs. Imprint:Prosiding seminar nasional analisis aktivasi neutron