CT colonography at low tube potential: using iterative reconstruction to decrease noise

  • 1. Department of Diagnostic Imaging, The Warren Alpert Medical School of Brown University, Rhode Island Hospital, 593 Eddy St., Providence, RI 02903 (United States)
  • 2. The Russell H. Morgan Department of Radiology and Radiological Science, The Johns Hopkins School of Medicine, 601 N. Caroline St, Room 3140D, Baltimore, MD 21287 (United States)

Description

Aim: To determine the level of iterative reconstruction required to reduce increased image noise associated with low tube potential computed tomography (CT). Materials and methods: Fifty patients underwent CT colonography with a supine scan at 120 kVp and a prone scan at 100 kVp with other scan parameters unchanged. Both scans were reconstructed with filtered back projection (FBP) and increasing levels of adaptive statistical iterative reconstruction (ASiR) at 30%, 60%, and 90%. Mean noise, soft tissue and tagged fluid attenuation, contrast, and contrast-to-noise ratio (CNR) were collected from reconstructions at both 120 and 100 kVp and compared using a generalised linear mixed model. Results: Decreasing tube potential from 120 to 100 kVp significantly increased image noise by 30–34% and tagged fluid attenuation by 120 HU at all ASiR levels (p<0.0001, all measures). Increasing ASiR from 0% (FBP) to 30%, 60%, and 90% resulted in significant decreases in noise and increases in CNR at both tube potentials (p<0.001, all comparisons). Compared to 120 kVp FBP, ASiR greater than 30% at 100 kVp yielded similar or lower image noise. Conclusions: Iterative reconstruction adequately compensates for increased image noise associated with low tube potential imaging while improving CNR. An ASiR level of approximately 50% at 100 kVp yields similar noise to 120 kVp without ASiR. -- Highlights: •Peak kilovoltage (kVp) can be reduced to decrease radiation dose and increase contrast attenuation at a cost of increased image noise. •Utilizing iterative reconstruction can decrease image noise and increase contrast to noise ratio (CNR) independent of kVp. •Iterative reconstruction adequately compensates for increased image noise associated with low dose low kVp imaging while improving CNR. •An ASiR level of approximately 50% at 100 kVp yields similar noise to 120 kVp without ASiR

Availability note (English)

Available from http://dx.doi.org/10.1016/j.crad.2015.05.005

Additional details

Identifiers

DOI
10.1016/j.crad.2015.05.005;
PII
S0009-9260(15)00170-1;

Publishing Information

Journal Title
Clinical Radiology
Journal Volume
70
Journal Issue
9
Journal Page Range
p. 981-988
ISSN
0009-9260
CODEN
CLRAAG

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
47025608
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ANIMAL TISSUES; ATTENUATION; BIOMEDICAL RADIOGRAPHY; COMPARATIVE EVALUATIONS; COMPUTERIZED TOMOGRAPHY; FILTERS; FLUIDS; IMAGES; ITERATIVE METHODS; NOISE; PATIENTS; RADIATION DOSES
Descriptors DEC
BODY; CALCULATION METHODS; DIAGNOSTIC TECHNIQUES; DOSES; EVALUATION; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; TOMOGRAPHY

Optional Information

Copyright
Copyright (c) 2015 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.