An extended range well counter

  • 1. Univ. of Wisconsin, Madison, WI

Description

A well-type detector is described for the assay of gamma emitters over a Curie-to-nanoCurie activity range. A large plastic scintillator/photomultiplier is DC-coupled for current measurements at high activities, with pulse counting covering the low activities. The two techniques overlap for two decases in the uCi regulation, particularly troublesome for the alternative detectors: gas-filled ionization chambers or NaI(T1) well counters. The plastic well is interfaced to a microcomputer for data acquistion and control of the phototube high voltage, the key to the nine-decase dynamic range

Additional details

Publishing Information

Journal Title
IEEE Trans. Nucl. Sci.
Journal Volume
NS-32
Journal Issue
1
Series
IEEE Trans. Nucl. Sci.
Journal Page Range
60-62
ISSN
0018-9499
CODEN
IETNA

Conference

Title
Nuclear science symposium.
Dates
31 Oct - 2 Nov 1984.
Place
Orlando, FL (USA).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
17051963
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
COMPUTERS; DATA ACQUISITION; GAMMA DETECTION; NAI DETECTORS; PHOTOMULTIPLIERS; PLASTIC SCINTILLATORS; RADIATION DETECTORS
Descriptors DEC
DETECTION; MEASURING INSTRUMENTS; PHOSPHORS; PHOTOTUBES; RADIATION DETECTION; SCINTILLATION COUNTERS; SOLID SCINTILLATION DETECTORS

Optional Information

Secondary number(s)
CONF-841007--.