Published December 7, 2015 | Version v1
Journal article

Automated measurement of CT noise in patient images with a novel structure coherence feature

  • 1. Interdisciplinary Program of Radiation Applied Life Science, Seoul National University College of Medicine, Seoul, 110–744 (Korea, Republic of)
  • 2. Department of Radiology, Seoul National University Hospital, Seoul, 156–756 (Korea, Republic of)

Description

While the assessment of CT noise constitutes an important task for the optimization of scan protocols in clinical routine, the majority of noise measurements in practice still rely on manual operation, hence limiting their efficiency and reliability. This study presents an algorithm for the automated measurement of CT noise in patient images with a novel structure coherence feature. The proposed algorithm consists of a four-step procedure including subcutaneous fat tissue selection, the calculation of structure coherence feature, the determination of homogeneous ROIs, and the estimation of the average noise level. In an evaluation with 94 CT scans (16 517 images) of pediatric and adult patients along with the participation of two radiologists, ROIs were placed on a homogeneous fat region at 99.46% accuracy, and the agreement of the automated noise measurements with the radiologists' reference noise measurements (PCC  =  0.86) was substantially higher than the within and between-rater agreements of noise measurements (PCCwithin  =  0.75, PCCbetween  =  0.70). In addition, the absolute noise level measurements matched closely the theoretical noise levels generated by a reduced-dose simulation technique. Our proposed algorithm has the potential to be used for examining the appropriateness of radiation dose and the image quality of CT protocols for research purposes as well as clinical routine. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/0031-9155/60/23/9107

Additional details

Identifiers

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
60
Journal Issue
23
Journal Page Range
p. 9107-9122
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
47075866
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ACCURACY; ALGORITHMS; ANIMAL TISSUES; FATS; IMAGE PROCESSING; IMAGES; OPTIMIZATION; PATIENTS; PEDIATRICS; RADIATION DOSES
Descriptors DEC
BODY; DOSES; MATHEMATICAL LOGIC; MEDICINE; PROCESSING