Investigation of percentage depth dose and dose rate dependence of PAGAT polymer gel dosimeter for photon beams using MRI technique
Creators
- 1. Borujerd Islamic Aad University, Faculty of Engineering, Borujerd (Iran, Islamic Republic of)
- 2. Shiraz University, Faculty of Engineering, Shiraz (Iran, Islamic Republic of)
- 3. Novin Medical Radiation Institute, Tehran (Iran, Islamic Republic of)
Description
In this work, the investigation of the normoxic Pagan polymer-gel dosimeter percentage depth dose and it's dose rate dependence has been made. Using MRI, the formulation to give the maximum change in the transverse relaxation rate R2 was determined to be 4.5% N,N'-methylene-bis-acrylamide(bis), 4.5% acrylamide, 5% gelatine, 5 m M tetrakis (hydroxymethyl) phosphonium chloride, 0.01 m M hydroquinone and 86% HPLC(Water). Irradiation of vials was performed using photon beams of Co-60 therapy unit and an Elec ta linear accelerator. Gel dosimeters were imaged in a Siemens Symphony 1.5 Tesla clinical MRI scanner using a head coil. Post-manufacture irradiation and post imaging times were both selected to be 1 day. For determining the percentage depth dose of the Pagan gel it was found that at the depth of 21 cm, the percentage depth dose for 1.25 MeV γ-ray photons of 60Co and for 4,6 and 18 MV x-ray photons of Elec ta linear accelerator, are 48%, 52%, 57.3% and 59.73%, respectively. Thus, in the case of the higher energy photon beams, a higher dose can he delivered. to deep-seated tumors. The dose rate dependence of percentage depth dose was studied for 6 MV x-ray photons with the use of dose rates of 80, 160, 240, 320, 400 and 480 c Gy/min. No trend in polymer-gel dosimeter 1/T 2 dependence was found on the mean dose rate and energy for the photon beams.
Availability note (English)
Available from Atomic Energy Organization of IranAdditional details
Additional titles
- Original title (Persian)
- Barresi-e darsad-e doz-e omghi-e barike-haye fotoni va vabastegi-e an be ahang-e doz dar dosimetr-e zheli-polimeri-e noormoksik-e PAGAT ba estefade az teknik-e MRI
Publishing Information
- Journal Title
- Journal of Nuclear Science and Technology
- Journal Issue
- no.51
- Journal Page Range
- p. 70-76
- ISSN
- 1735-1871
INIS
- Country of Publication
- Iran, Islamic Republic of
- Country of Input or Organization
- Iran, Islamic Republic of
- INIS RN
- 41128757
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ACRYLAMIDE; CHEMICAL DOSEMETERS; COBALT 60; DEPTH DOSE DISTRIBUTIONS; DOSE RATES; GAMMA RADIATION; GELS; IRRADIATION; LINEAR ACCELERATORS; NMR IMAGING; PHOTON BEAMS; RELAXATION; SENSITIVITY; X RADIATION
- Descriptors DEC
- ACCELERATORS; AMIDES; BEAMS; BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; COBALT ISOTOPES; COLLOIDS; DIAGNOSTIC TECHNIQUES; DISPERSIONS; DOSEMETERS; ELECTROMAGNETIC RADIATION; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTOPES; IONIZING RADIATIONS; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; MEASURING INSTRUMENTS; MINUTES LIVING RADIOISOTOPES; NUCLEI; ODD-ODD NUCLEI; ORGANIC COMPOUNDS; ORGANIC NITROGEN COMPOUNDS; RADIATION DOSE DISTRIBUTIONS; RADIATIONS; RADIOISOTOPES; SPATIAL DOSE DISTRIBUTIONS; YEARS LIVING RADIOISOTOPES