Published December 11, 1995 | Version v1
Journal article

Position-sensitive MCP-based detectors with high timing resolution: some results and perspectives

  • 1. Sankt Peterburgskij Gosudarstvennyj Univ., St. Petersburg (Russian Federation)

Description

The microelectronic design and some results of experimental tests of three types of coordinate-sensitive Micro-Channel-Plane (MCP) based detectors with good timing properties are presented. The electrodynamic structures that could be used for the MCP fast signal readout (strips, delay lines and discrete pads) are always a compromise between the amplitude of the induced signal, propagation time delay, cross-talk and distortions of the signal at the output. In-lab tests of these structures and experimental tests of the detectors done with the beams of MIPs and with the alpha-source, have shown a 130 μm spatial resolution and a 110 ps timing resolution obtained simultaneously for one detector. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
367
Journal Issue
1-3
Journal Page Range
p. 402-407.
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
7. international wire chamber conference.
Dates
13-17 Feb 1995.
Place
Vienna (Austria).