Quantification and isotope ratio measurement of boron in U3Si2 by inductively coupled plasma mass spectrometry

  • 1. Radioanalytical Chemistry Division, Bhabha Atomic Research Centre, Mumbai (India)
  • 2. Integral Fuel Fabrication Facility, Bhabha Atomic Research Centre, Mumbai (India)

Description

An analytical methodology was developed for precise quantification and isotope ratio measurement of boron in U3Si2 matrix by using ICP-MS after matrix separation. The analytical technique was validated by recovery studies employing standard addition method and the accuracy in isotope ratio measurement was improved by correcting the bias factor after analyzing NIST SRM951. The quantification of B in the three U3Si2 samples was found in the range of 2.32-3.90 μg g-1 with a maximum standard deviation of 3%. The 10B/11B value in the three samples was found to be 0.2455±0.0042, 0.2451±0.0036 and 0.2452±0.0041. (author)

Part of:
Proceedings of the DAE-BRNS fifth symposium on nuclear analytical chemistry

Additional details

Publishing Information

Publisher
Bhabha Atomic Research Centre
Imprint Place
Mumbai (India)
Imprint Title
Proceedings of the DAE-BRNS fifth symposium on nuclear analytical chemistry
Imprint Pagination
603 p.
Journal Page Range
p. 284-285

Conference

Title
5. DAE-BRNS symposium on nuclear analytical chemistry
Acronym
NAC-V
Dates
20-24 Jan 2014
Place
Mumbai (India)

INIS

Optional Information

Notes
7 refs., 2 tabs.