A method of predicting the image noise in paediatric multi-slice computed tomography images

  • 1. Dept. of Radiation Physics, Sahlgrenska Univ. Hospital, SE-413 45 Goeteborg (Sweden)
  • 2. Dept. of Paediatric Radiology and Physiology, The Queen Silvia Children's Hospital, SE-416 85 Goeteborg (Sweden)
  • 3. Dept. of Medical Physics and Biomedical Engineering, Sahlgrenska Univ. Hospital, SE-413 45 Goeteborg (Sweden)

Description

The purpose of this study was to develop an equation with which to determine the tube current to be used in order to obtain a certain image noise level for differently sized children undergoing multi-slice computed tomography examination. The relationship between image noise and detector dose for different examination protocols was established for a LightSpeed Ultra, an eight slice CT from GEMS, using homogeneous water phantoms of different sizes. Three different anatomical areas (head, thorax and abdomen) were studied in 111 patients between 0 and 17 y of age. The mean ratio between the calculated and the measured noise in patient images was established for the different areas. Head examinations showed the best correlation (measured-to-calculated noise ratio = 1.01). In the thorax, the calculated noise was generally higher than the measured noise (ratio = 0.74), and in the abdomen, the opposite result was found (ratio = 1.20). (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/nch568

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
114
Journal Issue
1-3
Journal Page Range
p. 313-316
ISSN
0144-8420

Conference

Title
2. Malmoe Conference on Medical X-Ray Imaging - Optimisation strategies in medical X-ray imaging
Dates
23-25 Apr 2004
Place
Malmoe (Sweden)

Optional Information

Notes
10 refs