Stability analysis of a deterministic dose calculation for MRI-guided radiotherapy
Creators
- 1. Department of Oncology, University of Alberta, 11560 University Ave, Edmonton, Alberta T6G 1Z2 (Canada)
Description
Modern effort in radiotherapy to address the challenges of tumor localization and motion has led to the development of MRI guided radiotherapy technologies. Accurate dose calculations must properly account for the effects of the MRI magnetic fields. Previous work has investigated the accuracy of a deterministic linear Boltzmann transport equation (LBTE) solver that includes magnetic field, but not the stability of the iterative solution method. In this work, we perform a stability analysis of this deterministic algorithm including an investigation of the convergence rate dependencies on the magnetic field, material density, energy, and anisotropy expansion. The iterative convergence rate of the continuous and discretized LBTE including magnetic fields is determined by analyzing the spectral radius using Fourier analysis for the stationary source iteration (SI) scheme. The spectral radius is calculated when the magnetic field is included (1) as a part of the iteration source, and (2) inside the streaming-collision operator. The non-stationary Krylov subspace solver GMRES is also investigated as a potential method to accelerate the iterative convergence, and an angular parallel computing methodology is investigated as a method to enhance the efficiency of the calculation. SI is found to be unstable when the magnetic field is part of the iteration source, but unconditionally stable when the magnetic field is included in the streaming-collision operator. The discretized LBTE with magnetic fields using a space-angle upwind stabilized discontinuous finite element method (DFEM) was also found to be unconditionally stable, but the spectral radius rapidly reaches unity for very low-density media and increasing magnetic field strengths indicating arbitrarily slow convergence rates. However, GMRES is shown to significantly accelerate the DFEM convergence rate showing only a weak dependence on the magnetic field. In addition, the use of an angular parallel computing strategy is shown to potentially increase the efficiency of the dose calculation. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/1361-6560/aa959aAdditional details
Identifiers
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 63
- Journal Issue
- 1
- Journal Page Range
- [16 p.]
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 52002711
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BOLTZMANN EQUATION; CONVERGENCE; FINITE ELEMENT METHOD; ITERATIVE METHODS; MAGNETIC FIELDS; NMR IMAGING; RADIATION DOSES; RADIOTHERAPY
- Descriptors DEC
- CALCULATION METHODS; DIAGNOSTIC TECHNIQUES; DIFFERENTIAL EQUATIONS; DOSES; EQUATIONS; INTEGRO-DIFFERENTIAL EQUATIONS; KINETIC EQUATIONS; MATHEMATICAL SOLUTIONS; MEDICINE; NUCLEAR MEDICINE; NUMERICAL SOLUTION; PARTIAL DIFFERENTIAL EQUATIONS; RADIOLOGY; THERAPY