Influence of some instrumental parameters on the determination of quantitative data using the PET scanner PC4096-15WB

  • 1. Kernforschungsanlage Juelich G.m.b.H. (Germany, F.R.). Inst. fuer Medizin

Description

In order to evaluate the influence of physical and instrumental limitations of the PET technique on metabolic data, various parameters have been tested. The measurements were done with a whole body PET scanner PC4096-15WB. The dependence on the transaxial resolution of reconstruction filter cut off frequency αc was investigated. The high resolution (4.8 mm at r=0 in the field of view) and the low scatter fraction (<5%) of this system yields a higher grey/white matter ratio (3.8:1) than that found with older scanners (2:1). The use of a rotating pin source allows transmission scans to be performed after tracer administration with an error of less than 5%. (orig.)