Published 1976
| Version v1
Book
Drain feedback preamplifier for low energy X-ray spectroscopy work
- 1. Bhabha Atomic Research Centre, Bombay (India). Electronics Div.
- 2. Bhabha Atomic Research Centre, Bombay (India). Nuclear Physics Div.
Description
A new method of charge restoration in low noise preamplifiers has been described. Using six microsecond shaping, with a small area 3 mm SiLi, a resolution (fwhm) of 280 eV was achieved for the manganese Kα peak. (author)
Additional details
Publishing Information
- Publisher
- Department of Atomic Energy.
- Imprint Place
- Bombay
- Imprint Title
- Proceedings of the nuclear physics and solid state physics symposium, Calcutta, December 22-26, 1975
- Imprint Pagination
- p. 329.
Conference
- Title
- Nuclear physics and solid state physics symposium.
- Dates
- 22 - 26 Dec 1975.
- Place
- Calcutta, India.
INIS
- Country of Publication
- India
- Country of Input or Organization
- India
- INIS RN
- 8318854
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- EV RANGE 01-10; EV RANGE 10-100; PREAMPLIFIERS; RESOLUTION; X-RAY SPECTROMETERS
- Descriptors DEC
- AMPLIFIERS; ELECTRONIC EQUIPMENT; ENERGY RANGE; EV RANGE; MEASURING INSTRUMENTS; SPECTROMETERS