Published May 2017
| Version v1
Journal article
Patient size based guiding equations for automatic mAs and kVp selections in general medical X-ray projection radiography
Creators
- 1. Medical Radiation Science, School of Dentistry and Health Sciences, Faculty of Science, Charles Sturt University, Wagga Wagga, New South Wales 2678 (Australia)
Description
A patient size based guiding equation for the automatic selections of mAs and peak voltage kVp in general medical X-ray projection radiography was derived from the first principles of dose and image quality optimization. Under various specific conditions of constant patient size d, kVp or mAs, this equation leads to various longstanding 'rules of thumbs' currently being employed in clinical practice. For automatic mAs control, this work suggests that the current level of dose to patient in X-ray radiography should be halved without compromise image quality. Further studies on the dependence of the absorbed dose on the patient's thickness are required in general X-ray projection radiography. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1093/rpd/ncw246Additional details
Identifiers
- DOI
- 10.1093/rpd/ncw246;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 174
- Journal Issue
- 4
- Journal Page Range
- p. 545-550
- ISSN
- 0144-8420
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 48063661
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ABSORBED RADIATION DOSES; ELECTRIC POTENTIAL; EQUATIONS; IMAGES; OPTIMIZATION; PATIENTS; THICKNESS; X-RAY RADIOGRAPHY
- Descriptors DEC
- DIMENSIONS; DOSES; INDUSTRIAL RADIOGRAPHY; MATERIALS TESTING; NONDESTRUCTIVE TESTING; RADIATION DOSES; TESTING
Optional Information
- Notes
- 26 refs.