AP-002M analog processor for ionizing radiation spectrometers with semiconductor detector

  • 1. Ural'skij Politekhnicheskij Inst., Sverdlovsk (USSR)

Description

A fast-responce unit of complex analog processing of spectrometric pulses from gamma and X-ray semiconductor detectors is described. Analog processor (AP) AP-002M contains: a linear amplifier with galvanic-connected amplification sections; spectrometric time-variant shaper involving delay line differentiator, time-variant initial level stabilizer and strobing ''ideal'' integrator; overload pulse rejector; superposed pulse rejector; expander-threshold device; sretcher; AP indication state auxiliary devices. The gain varies in the range 16-1024, shaping time is 2-12 μs. Maximal statistic loading at AP-002M input attains 105s-1 for all shaping times. Analog processor output loading is approximately three times as much that for optimized with respect to noise Gaussian amplifier with superposition rejector. The peak gravity centre shift Ex=5.9 keV is not over 0.15%

Additional details

Additional titles

Original title (Russian)
Аналоговый процессор AP-002M для спеcтрометров ионизирующих излучений с полупроводниковым детеcтором

Publishing Information

Journal Title
Prib. Tekh. Ehksp.
Journal Issue
no.5
Series
Prib. Tekh. Ehksp.
ISSN
0032-8162
CODEN
PRTEA