Published October 1, 2011 | Version v1
Journal article

Dose-rate and the reciprocity law: TL response of Ge-doped SiO2 optical fibers at therapeutic radiation doses

  • 1. School of Physics and Material Studies, Faculty of Applied Sciences, Universiti Teknologi MARA Malaysia (UiTM), Campus of Negeri Sembilan, 72000 Kuala Pilah (Malaysia)
  • 2. Centre for Nuclear and Radiation Physics, Department of Physics, University of Surrey, Guildford GU2 7XH, Surrey (United Kingdom)
  • 3. Departments of Medical Physics, the Royal Surrey County Hospital (RSCH) NHS Trust, Edgerton Road, Guildford GU2 7XX, Surrey (United Kingdom)

Description

An investigation has been made on commercially available Ge-doped SiO2 optical fibers as a novel thermoluminescence system for radiotherapy dosimetry. This dosimeter has previously been shown by the group to provide sensitive dosimetry over a wide range of electron and photon dose, suitable for the needs of radiotherapy. In addition the optical fiber offers small physical size (125 μm diameter) and hence high spatial resolution. The reciprocity between thermoluminescence (TL) yield of Ge-doped SiO2 optical fibers and dose has been investigated for fixed radiation dose for a range of photon and electron dose rates. For electron beams of nominal energies in the range of 9-20 MeV, we have investigated the TL response of these fibers for dose rates between 100 and 1000 cGy min-1. For photon beams of nominal energies in the range of 6-15 MV, we have used dose rates of 100-600 cGy min-1. Reproducibility and fading at fixed absorbed dose (3 Gy) and dose rate for the optical fibers were also investigated. At fixed dose rates, the TL optical fibers were found to produce a flat TL yield within 4% (1σ) and 3% (1σ) for electron and photon beams, respectively. The optical fibers demonstrated good reproducibility (±1.5%), low residual signal for a readout temperature of 300 oC and negligible fading. A weak dependence on dose-rate has been observed in the range of 3.4-3.9% for electrons (with an associated uncertainty of 4%) and 2.4-2.9% for photons (with an associated uncertainty of <4%). For electron and photon energies we note a consistent trend towards lower response in the TL yield of between 3.4-3.9% and 2.4-2.7%, respectively, at the higher dose rates in comparison with the response at lower dose rates. In addition we note an appreciable systematic energy dependence for both electron and photon beams. It is important to take such factors into account for providing precise and accurate radiotherapy dosimetry. It is also apparent that the optical fibers can be re-used following thermal annealing, without being detrimental to dose response. Various features make these commercial optical fibers a promising TL material for use as a dosimetric system in radiotherapy.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2010.08.048

Additional details

Identifiers

DOI
10.1016/j.nima.2010.08.048;
PII
S0168-9002(10)01822-X;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
652
Journal Issue
1
Journal Page Range
p. 891-895
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
12. Symposium on radiation measurements and applications (SORMA)
Dates
24-28 May 2010
Place
Ann Arbor, MI (United States)

Optional Information

Copyright
Copyright (c) 2010 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.