Published 2016 | Version v1
Journal article

Effect of Radionuclide Activity Concentration on PET-CT Image Uniformity

  • 1. Medical Radiation Physics Centre, Radiological and Medical Sciences Research Institute, Ghana Atomic Energy Commission, Accra (Ghana)
  • 2. Department of Medical Physics, School of Nuclear and Allied Sciences, University of Ghana, Accra (Ghana)
  • 3. Department of Nuclear Medicine, CM Johannesburg Academic Hospital, University of the Witwatersrand, Johannesburg (South Africa)

Description

Assessment of radionuclide activity concentration on positron emission tomography-computedr tomography (PET-CT) image uniformity has been carried out quantitatively. Tomographic PET-CT images of cylindrical phantom containing F-18 fluorodeoxyglucose (FDG) activity concentration was acquired and used for the assessment. Activity concentrations were varied and PET-CT images were acquired at the constant acquisition parameters of time, matrix size, and reconstruction algorithm, respectively. Using midtransaxial image slices, quantitative index of nonuniformity (NU), and coefficient of uniformity variation were estimated for the different activity concentrations. Maximum NUs of 17.6%, 26.3%, 32.7%, 36.2%, and 38.5% were estimated for activity concentrations of 16.87 kBq/mL, 14.06 kBq/mL, 11.25 kBq/mL, 8.43 kBq/mL, and 5.62 kBq/mL, respectively. The coefficient of uniformity variation established an inverse quadratic relationship with activity concentration. Activity concentrations of 16.87 kBq/mL, 14.06 kBq/mL, 11.25 kBq/mL, 8.43 kBq/mL, and 5.62 kBq/mL produced uniformity variations of 1.47%, 2.52%, 4.23%, 5.12%, and 4.98%, respectively. Increasing activity concentration resulted in decreasing coefficient of uniformity and hence, an increase in image uniformity. The uniformity estimates compared well with the standards set internationally

Availability note (English)

Available from http://dx.doi.org/10.4103/1450-1147.167578; Available from http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4809161

Additional details

Publishing Information

Journal Title
World Journal of Nuclear Medicine (Online)
Journal Volume
15
Journal Issue
2
Journal Page Range
p. 91-95
ISSN
1607-3312

Optional Information

Notes
PMCID: PMC4809161; PMID: 27134558; PUBLISHER-ID: WJNM-15-91; OAI: oai:pubmedcentral.nih.gov:4809161; Copyright: (c) World Journal of Nuclear Medicine; This record replaces 47118587