Description
The aim of this study was to find the optimum tube voltage for neonatal chest imaging in computed radiography. The study was designed to take full advantage of the benefits of digital imaging, for example, by comparing the tube voltages at constant effective dose. A phantom study using a living rabbit was first conducted. Images were collected at tube voltages ranging from 40 to 90 kVp. The reproduction of four structures (central vessels, peripheral vessels, carina and thoracic vertebrae) was rated by 10 radiologist. The reproduction of both central and peripheral vessels was relatively independent of tube voltage. The carina was better reproduced at higher tube voltages whereas the opposite was true for the thoracic vertebrae. Based on the higher importance of the reproduction of the carina it was decided that 90 kVp was the optimal tube voltage. To validate the result from the phantom study, a follow-up study was conducted in which images of neonates collected at the tube voltage regularly used at Sahlgrenska Univ. Hospital (70 kVp) were compared with images collected at the tube voltage proposed by the phantom study. The follow-up study confirmed the results from the phantom study that the reproduction of the carina was better at 90 than at 70 kVp. In conclusion, for neonatal chest imaging - given the same effective dose - 90 kVp gives better reproduction of important structures than the regularly used 70 kVp. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1093/rpd/nch528Additional details
Identifiers
- DOI
- 10.1093/rpd/nch528;
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 114
- Journal Issue
- 1-3
- Journal Page Range
- p. 278-285
- ISSN
- 0144-8420
Conference
- Title
- 2. Malmoe Conference on Medical X-Ray Imaging - Optimisation strategies in medical X-ray imaging
- Dates
- 23-25 Apr 2004
- Place
- Malmoe (Sweden)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- France
- INIS RN
- 44035807
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE; S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CHEST; COMPUTERIZED TOMOGRAPHY; HOSPITALS; IMAGE PROCESSING; IMAGES; INFANTS; OPTIMIZATION; PHANTOMS; RABBITS; RADIATION DOSES; RADIOLOGY; VERTEBRAE; X RADIATION
- Descriptors DEC
- AGE GROUPS; ANIMALS; BODY; BUILDINGS; CHILDREN; DIAGNOSTIC TECHNIQUES; DOSES; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MAMMALS; MAN; MEDICAL ESTABLISHMENTS; MEDICINE; MOCKUP; NUCLEAR MEDICINE; ORGANS; PRIMATES; PROCESSING; RADIATIONS; SKELETON; STRUCTURAL MODELS; TOMOGRAPHY; VERTEBRATES
Optional Information
- Notes
- 15 refs