Dosimetric methodology and reference system for diagnostic level X radiation

Description

Several methodologies for the calibration of diagnostic radiology instruments were developed and established at the Calibration Laboratory of IPEN. These established radiation qualities are recommended by international standards. The methods may be used in the calibration procedures of survey meters used in radiation protection measurements (scattered radiation), instruments used in direct beams (attenuated and non attenuated beams) and quality control instruments. A reference system was proposed using two identical ionization chambers developed at IPEN. They differ only by the collecting electrode material, one of aluminium and the other of graphite. The different energetic dependence of the chamber's response provided a ratio related to the tube potential. The variation of only 0.28%, from 14.3 to 111 keV, on the energetic dependence of the graphite electrode chamber, provided the possibility of air kerma rate determination in the studied radiation beams. (author)