Inhomogeneity effect in Varian Trilogy Clinac iX 10 MV photon beam using EGSnrc and Geant4 code system

  • 1. Department of Physics, Faculty of Mathematics and Natural Sciences, Institut Teknologi Bandung, Jalan Ganesa 10 Bandung 40132 (Indonesia)
  • 2. Department of Radiation Oncology, National University Cancer Institute Singapore, Singapore/Singapore (Singapore)

Description

Treatment fields consist of tissue other than water equivalent tissue (soft tissue, bones, lungs, etc.). The inhomogeneity effect can be investigated by Monte Carlo (MC) simulation. MC simulation of the radiation transport in an absorbing medium is the most accurate method for dose calculation in radiotherapy. The aim of this work is to evaluate the effect of inhomogeneity phantom on dose calculations in photon beam radiotherapy obtained by different MC codes. MC code system EGSnrc and Geant4 was used in this study. Inhomogeneity phantom dimension is 39.5 × 30.5 × 30 cm3 and made of 4 material slices (12.5 cm water, 10 cm aluminium, 5 cm lung and 12.5 cm water). Simulations were performed for field size 4 × 4 cm2 at SSD 100 cm. The spectrum distribution Varian Trilogy Clinac iX 10 MV was used. Percent depth dose (PDD) and dose profile was investigated in this research. The effects of inhomogeneities on radiation dose distributions depend on the amount, density and atomic number of the inhomogeneity, as well as on the quality of the photon beam. Good agreement between dose distribution from EGSnrc and Geant4 code system in inhomogeneity phantom was observed, with dose differences around 5% and 7% for depth doses and dose profiles. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1742-6596/739/1/012117

Additional details

Publishing Information

Journal Title
Journal of Physics. Conference Series (Online)
Journal Volume
739
Journal Issue
1
Journal Page Range
[7 p.]
ISSN
1742-6596

Conference

Title
6. Asian physics symposium
Dates
19-20 Aug 2015
Place