Published January 2003
| Version v1
Journal article
Performance evaluation of thermoluminescent cards using nitrogen generator in comparison with commercial nitrogen and compressed air
- 1. Health and Safety Division, Indira Gandhi Centre for Atomic Research, Kalpakkam (India)
Description
Performance of the TLD cards is evaluated from the dose point of view by heating them under three different conditions viz. nitrogen from nitrogen generator, nitrogen from commercial nitrogen cylinders and compressed air cylinder and the results are compared with true dose. To begin with, the background reading and its standard deviation are found to be more for those cards read by using compressed air compared to nitrogen generator. About 100 cards were used for comparison for the first cycle. Out of this, a total of 30 cards were subjected to 10 test cycles. It is observed that the performance of the cards before and after 10 test cycles does not vary much. The maximum deviation observed is about ±8% from the true dose. (author)
Additional details
Publishing Information
- Journal Title
- Radiation Protection and Environment
- Journal Volume
- 26
- Journal Issue
- 1-2,pt.2
- Journal Page Range
- p. 314-316
- CODEN
- RPREFM
Conference
- Title
- 26. IARP conference on radiation exposure control at nuclear fuel cycle facilities and radiation installations
- Acronym
- IARP-NC.2K3
- Dates
- 5-7 Mar 2003
- Place
- Kalpakkam (India)
INIS
- Country of Publication
- India
- Country of Input or Organization
- India
- INIS RN
- 35072992
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- GAS GENERATORS; NITROGEN; OCCUPATIONAL EXPOSURE; PERFORMANCE; PERSONNEL MONITORING; RADIATION DOSES; RADIATION PROTECTION; THERMOLUMINESCENT DOSEMETERS
- Descriptors DEC
- DOSEMETERS; DOSES; ELEMENTS; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS; MONITORING; NONMETALS; RADIATION MONITORING
Optional Information
- Notes
- 3 tabs.