Performance evaluation of thermoluminescent cards using nitrogen generator in comparison with commercial nitrogen and compressed air

  • 1. Health and Safety Division, Indira Gandhi Centre for Atomic Research, Kalpakkam (India)

Description

Performance of the TLD cards is evaluated from the dose point of view by heating them under three different conditions viz. nitrogen from nitrogen generator, nitrogen from commercial nitrogen cylinders and compressed air cylinder and the results are compared with true dose. To begin with, the background reading and its standard deviation are found to be more for those cards read by using compressed air compared to nitrogen generator. About 100 cards were used for comparison for the first cycle. Out of this, a total of 30 cards were subjected to 10 test cycles. It is observed that the performance of the cards before and after 10 test cycles does not vary much. The maximum deviation observed is about ±8% from the true dose. (author)

Additional details

Publishing Information

Journal Title
Radiation Protection and Environment
Journal Volume
26
Journal Issue
1-2,pt.2
Journal Page Range
p. 314-316
CODEN
RPREFM

Conference

Title
26. IARP conference on radiation exposure control at nuclear fuel cycle facilities and radiation installations
Acronym
IARP-NC.2K3
Dates
5-7 Mar 2003
Place
Kalpakkam (India)

INIS

Country of Publication
India
Country of Input or Organization
India
INIS RN
35072992
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
GAS GENERATORS; NITROGEN; OCCUPATIONAL EXPOSURE; PERFORMANCE; PERSONNEL MONITORING; RADIATION DOSES; RADIATION PROTECTION; THERMOLUMINESCENT DOSEMETERS
Descriptors DEC
DOSEMETERS; DOSES; ELEMENTS; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS; MONITORING; NONMETALS; RADIATION MONITORING

Optional Information

Notes
3 tabs.