Published January 15, 1990 | Version v1
Journal article

Precise determination of gamma-ray peak areas

  • 1. Physikalisch-Technische Bundesanstalt, Braunschweig (Germany, F.R.)

Description

Experimental pulse height distributions of ''gamma lines'' taken with a HPGe (high-purity germanium) detector are considered as the sum of exponentials convoluted with a Gaussian. With this model, peak areas have been evaluated from the parameters of the exponentials at gamma-ray energies ranging from 0.1 to 1.8 MeV. In measurements with 137Cs samples, the ratio of peak areas at 662 keV agreed with the given ratio of photon flux values to within ±0.2%. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research, Section A
Journal Volume
286
Journal Issue
3
Series
Nucl. Instrum. Methods Phys. Res., Sect. A.
Journal Page Range
415-420
ISSN
0168-9002
CODEN
NIMAE

Conference

Title
Spectrometric methods, measurements and evaluations in conjunction with the 10. session and the 12. biennial meeting of the International Committee for Radionuclide Metrology (ICRM).
Acronym
International symposium on nuclear decay data
Dates
6-8 Jun 1989.
Place
Braunschweig (Germany, F.R.).