Published September 23, 2011 | Version v1
Journal article

Liquid Xenon Detectors for Positron Emission Tomography

  • 1. TRIUMF, 4004 Wesbrook Mall, Vancouver, BC, V6T 2A3 (Canada)
  • 2. BC Cancer Agency, 600 West 10th Avenue, Vancouver, BC, V5Z 4E6 (Canada)
  • 3. Department of Physics and Astronomy, University of British Columbia, 6224 Agricultural Road, Vancouver, BC, V6T 1Z1 (Canada)
  • 4. University of Montreal, CP 6128 Succursale Centre-Ville, Montreal, Quebec, H3C 3J7 (Canada)

Description

PET is a functional imaging technique based on detection of annihilation photons following beta decay producing positrons. In this paper, we present the concept of a new PET system for preclinical applications consisting of a ring of twelve time projection chambers filled with liquid xenon viewed by avalanche photodiodes. Simultaneous measurement of ionization charge and scintillation light leads to a significant improvement to spatial resolution, image quality, and sensitivity. Simulated performance shows that an energy resolution of < 10% (FWHM) and a sensitivity of 15% are achievable. First tests with a prototype TPC indicate position resolution < 1 mm (FWHM).

Availability note (English)

Available from http://dx.doi.org/10.1088/1742-6596/312/6/062006

Additional details

Publishing Information

Journal Title
Journal of Physics. Conference Series (Online)
Journal Volume
312
Journal Issue
6
Journal Page Range
[6 p.]
ISSN
1742-6596

Conference

Title
International nuclear physics conference 2010
Acronym
INPC2010
Dates
4-9 Jul 2010
Place
Vancouver, BC (Canada)