Mobile 3D tomograph
- 1. Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin (Germany)
- 2. RayScan Technologies, Meersburg (Germany)
Description
Mobile tomographs often have the problem that high spatial resolution is impossible owing to the position or setup of the tomograph. While the tree tomograph developed by Messrs. Isotopenforschung Dr. Sauerwein GmbH worked well in practice, it is no longer used as the spatial resolution and measuring time are insufficient for many modern applications. The paper shows that the mechanical base of the method is sufficient for 3D CT measurements with modern detectors and X-ray tubes. CT measurements with very good statistics take less than 10 min. This means that mobile systems can be used, e.g. in examinations of non-transportable cultural objects or monuments. Enhancement of the spatial resolution of mobile tomographs capable of measuring in any position is made difficult by the fact that the tomograph has moving parts and will therefore have weight shifts. With the aid of tomographies whose spatial resolution is far higher than the mechanical accuracy, a correction method is presented for direct integration of the Feldkamp algorithm
Abstract (German)
Bei einem mobilen Aufbau eines Tomographen, ergibt sich oft das Problem, dass Standort oder Aufbau des Geraets eine hohe Ortsaufloesung nicht zulas-sen. Obwohl sich der von der Fa. Isotopenforschung Dr. Sauerwein GmbH entwi-ckelte Baumtomograph in der Praxis bewaehrte, kommt er heute nicht mehr zum Einsatz, da die zu erreichende Ortsaufloesung und die Messzeit fuer viele Fragestellun-gen heute als nicht mehr zeitgemaess erscheint. Wir zeigen hier, dass die mechanische Grundlage ausreicht, um 3D-CT Messungen mit neuen Detektoren und Roentgenroeh-ren durchzufuehren. CT Messungen mit sehr guter Statistik benoetigen weniger als 10min. Untersuchungen z. B. an nicht transportablen Kulturguetern sind so vor Ort moeglich. Einer darueber hinaus gehenden Erhoehung der Ortsaufloesung bei einem mobilen Tomographen, der in jeder Raumlage messen kann, steht jedoch eine Bewegung relativ zum Objekt, auf Grund der sich bewegenden Teile des Tomographen und der damit verbundenen Gewichtsverlagerung, entgegen. Anhand von Tomographien, deren Ortsaufloesung weit oberhalb der mechanischen Genauigkeit liegt, wird ein Korrek-turverfahren zur direkten Integration in den Feldkampalgorithmus vorgestelltAdditional details
Additional titles
- Original title (German)
- Mobiler 3D Tomograph
Publishing Information
- ISBN
- 978-3-940283-09-2
- Imprint Title
- 2008 annual meeting of the Deutsche Gesellschaft fuer Zerstoerungsfreie Pruefung (DGZfP): NDE in research, development and application
- Imprint Pagination
- 873 p.
- Journal Volume
- 113-CD
- Series
- Deutsche Gesellschaft fuer Zerstoerungsfreie Pruefung. Berichtsband
- Journal Page Range
- 7 p.
Conference
- Title
- NDE in research, development and application
- Original Conference Title
- Jahrestagung 2008 der Deutschen Gesellschaft fuer Zerstoerungsfreie Pruefung e.V. (DGZfP): ZfP in Forschung, Entwicklung und Anwendung
- Acronym
- 2008 annual meeting of the Deutsche Gesellschaft fuer Zerstoerungsfreie Pruefung (DGZfP)
- Dates
- 28-30 Apr 2008
- Place
- St. Gallen (Switzerland)
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 40015901
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference, Non-conventional Literature
- Descriptors DEI
- COMPUTERIZED TOMOGRAPHY; IMAGE PROCESSING; PORTABLE EQUIPMENT; SPATIAL RESOLUTION; X-RAY DETECTION
- Descriptors DEC
- DETECTION; DIAGNOSTIC TECHNIQUES; EQUIPMENT; PROCESSING; RADIATION DETECTION; RESOLUTION; TOMOGRAPHY
Optional Information
- Notes
- Imprint:DGZfP-Jahrestagung 2008. Zerstoerungsfreie Materialpruefung. ZfP in Forschung, Entwicklung und Anwendung