Evaluation of the linearity characteristic of the cone-beam CT fixed on the Varian 23EX linear accelerator
Creators
- 1. Department of Radiation Oncology, Zhongnan Hospital of Wuhan University (China)
Description
Objective: To investigate the CT number linearity of the cone-beam CT (CBCT) images at the different spatial locations in the scanning area. Methods: The Catphan 504 phantom at the different locations are scanned repeatedly using the CBCT on the Varian 23EX linear accelerator. The phantom is located the isocenter point, eccentric 3 cm, eccentric 6 cm, and different points on the z-axis successively on the accelerator. The scanned mode is the standard head mode. The reconstructive thickness is 2.5 cm. The different densities inserts of CTP 4.4 module on the different locations are measured via Eclips treatment planning system (TPS) and computed by Matlab 7.0 and the CT linear fitting are then processed. In order to understand better the linear distribution along with the value of CT in the spatial distribution the results are compared with the fan-beam CT. Results: Phantom studies show that: CBCT has good linearity performance not only under the standard header (body) of the scanning conditions, but also on such locations including the cross-sectional, the sagittal, the coronal plane and the eccentric position (R2>0.953). Bowtie filtration device dose not change the CT linearity but changes the value of CT. Conclusions: The linearity of X-ray CBCT on the Varian linear accelerator is favorable. CBCT will be used in the TPS dose calculation via further correction of the CT value. (authors)
Additional details
Publishing Information
- Journal Title
- Chinese Journal of Radiation Oncology
- Journal Volume
- 21
- Journal Issue
- 5
- Journal Page Range
- p. 464-467
- ISSN
- 1004-4221
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 45021481
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BEAMS; CAT SCANNING; CONES; CORRECTIONS; DIAGNOSIS; EVALUATION; IMAGE PROCESSING; IMAGES; LINEAR ACCELERATORS; PHANTOMS; RADIATION DOSES; SPATIAL DISTRIBUTION; X RADIATION
- Descriptors DEC
- ACCELERATORS; COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; DISTRIBUTION; DOSES; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MOCKUP; PROCESSING; RADIATIONS; STRUCTURAL MODELS; TOMOGRAPHY
Optional Information
- Notes
- 4 figs., 1 tabs., 9 refs.